One Step Cotinine Test Device
A rapid, one step test for the qualitative detection of Cotinine (nicotine
metabolite) in human urine.
For determination of Smoking Status only.
Intended Use
The COT one step Cotinine test device (urine) is a lateral flow
chromatographic immunoassay for the detection of Cotinine in human urine at a
cut-off concentration of 200 ng/mL.
This assay provides only a preliminary analytical result. A more specific
alternate chemical method must be used in order to obtain a confirmed analytical
result. Gas chromatography and mass spectrometry (GC/MS) is the preferred
confirmatory method. Clinical consideration and professional judgment should be
applied to any drug of abuse test result, particularly when preliminary positive
results are used.
Materials Provided
- Test device
- Dropper
- Package insert
Materials required but not provided
- Specimen Collection Container
- Timer
Directions for Use
The test device should remain in the sealed pouch until ready to use.
Allow test device, urine specimen to reach room temperature (15-30°C) prior
to testing.
- Bring pouch to room temperature before opening it. Remove the test
device from the sealed pouch and use it as soon as possible.
- Place the test device on a clean and level surface. Hold dropper
vertically and transfer 3 full drops of urine (approx. 100 uL) to the
specimen well (S) of the test device, and then start the timer. Avoid
trapping air bubbles in the specimen well (S). See illustration below.
- Wait for the colored line(s) to appear. The result should be read at
5 minutes. It is important that the background is clear before the
result is read. Do not interpret the result after 10 minutes.

Interpretation of Results
(Please refer to the illustration above)
Negative:* Two lines appear. One colored line should be in the control
line region ©, and another apparent colored line should be in the test line
region (T). This negative result indicates that the Cotinine concentration is
below the detectable level (200 ng/mL).
*Note: the shade of color in the test line region (T) may vary, but it
should be considered negative whenever there is even a faint line.
Positive: One colored line appears in the control line region (C). No
line appears in the test line region (T). This positive result indicates that
the Cotinine concentration exceeds the detectable level (200 ng/mL).
Invalid: Control line fails to appear. Insufficient specimen volume or
incorrect procedural techniques are the most likely for control line failure.
Review the procedure and repeat the test using a new test device. If the problem
persists, discontinue using the lot immediately and contact your local
distributor.